Magnetic field at
the supply fre-
quency (50/60
Hz)
IEC 61000-4-8
30 A/m 30 A/m Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environ-
ment.
Note: U
T
is the alternating current in the network before the application
of the test level.
Tests of Resist-
ance to Disrup-
tion
IEC 60601-1-2:2014
Test Level
Level of
Agreement
Electromagnetic
Environment
Conducted RF
IEC 61000-4-6
3 V
150 kHz to 80 MHz
ISM frequency band
Amateur radio band
3 V
See next ta-
ble
Portable and mo-
bile RF communi-
cations equip-
ment should be
used no closer to
any part of the
device, including
cables, than the
recommended
separation dis-
tance calculated
from the equa-
tion applicable to
the frequency of
the transmitter.
Recommended
protective dis-
tance:
d = 1.2
150
kHz to 80 MHz
d = 1.2 80
MHz to 800 MHz
d = 2.3
800
MHz to 2.5 GHz
where ’P’ is the
maximum output
power rating of
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
Proximity fields from
RF
10 V/m
See next ta-
ble
128 | DR 10e, DR 14e, DR 17e | Remarks for HF-emission and immunity