Table 12: Test compliance levels between portable and mobile RF
communications equipment and the DR detector
ISM (Industrial, Scientific and Medical) frequency band compliance lev-
el
Frequency
MHz
Test level
V
Frequency
MHz
Test level
V
6.765 6 13.553 6
26.957 6 40.66 6
Amateur radio band compliance level
Frequency
MHz
Test Level
V
Frequency
MHz
Test level
V
1.8 6 3.5 6
5.3 6 7 6
10.1 6 14 6
18.07 6 21 6
24.89 6 28 6
50 6
Proximity fields from RF compliance level
Frequency
MHz
Test level
V/m
Frequency
MHz
Test level
V/m
385 27 450 28
710 9 745 9
780 9 810 28
870 28 930 28
1462 10 1720 28
1845 28 1970 28
2450 28 3540 10
5240 9 5500 9
5785 9
130 | DR 10e, DR 14e, DR 17e | Remarks for HF-emission and immunity
0370C EN 20200228 1449