10 Agilent 1260 Infinity DAD and MWD User Manual
1
Introduction  
Introduction to the Detector
Introduction to the Detector
The detector is designed for highest optical performance, GLP compliance and 
easy maintenance. It includes the following features:
• 80 Hz data acquisition rate for (ultra-) fast LC applications (requires 
internal hard disk, G1315C and G1365C only),
• data recovery (DRC) feature provides data-never-lost insurance (requires 
internal hard disk, G1315C and G1365C only),
• RFID tags for all flow cells and UV-lamps provides traceable information 
about these assemblies, 
• long-life deuterium with RFID tag and tungsten lamps for highest intensity 
and lowest detection limit over a wavelength range of 190 – 950 nm,
• no loss in sensitivity for up to eight wavelengths simultaneous,
• programmable slit from 1 – 16 nm for complete optimization of sensitivity, 
linearity and spectral resolution,
• optional flow-cell cartridges with RFID tag (standard 10 mm13 μL, 
semi-micro 6 mm5 μL, micro 3 mm2 μL, 80 nL, 500 nL, 10 mm, high 
pressure 10 mm1.7 μL and prep-cells) are available and can be used 
depending on the application needs,
• easy front access to lamps and flow cell for fast replacement, and
• built-in holmium oxide filter for fast wavelength accuracy verification,
• built-in temperature control for improved baseline stability,
• additional diagnostic signals for temperature and lamp voltage monitoring,
For specifications, see “Performance Specifications” on page 24.