Agilent 1260 Infinity DAD and MWD User Manual 135
Error Information
7
Detector Error Messages
Diode Current Leakage
Error ID: 1041
When the detector is switched on, the processor checks the leakage current of
each of the optical diodes. If the leakage current exceeds the upper limit, the
error message is generated.
UV Lamp Current
Error ID: 7450
The UV lamp current is missing.
The processor continually monitors the anode current drawn by the lamp
during operation. If the anode current falls below the lower current limit, the
error message is generated.
Probable cause Suggested actions
1
Defective PDA/optical unit.
Please contact your Agilent service
representative.
2
Defective connector or cable.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Lamp disconnected.
Ensure the UV lamp connector is seated firmly.
2
Defective UV lamp or non-Agilent lamp.
Exchange the UV lamp.
3
Defective detector main board.
Please contact your Agilent service
representative.
4
Defective power supply.
Please contact your Agilent service
representative.