Agilent 1260 Infinity DAD and MWD User Manual 169
Maintenance
9
Exchanging a Lamp
Exchanging a Lamp
When If noise or drift exceeds application limits or lamp does not ignite
Tools required Description
Screwdriver, Pozidriv #1 PT3
Parts required # p/n Description
1 2140-0820 Longlife Deuterium lamp “C“ (with black cover and RFID tag)
1 G1103-60001 Tungsten lamp
Preparations Turn th e lamp(s) off.
WARNING
Eye damage by detector light
Eye damage may result from directly viewing the light produced by the deuterium
lamp used in this product.
Always turn the deuterium lamp off before removing it.
WARNING
Injury by touching hot lamp
If the detector has been in use, the lamp may be hot.
If so, wait for lamp to cool down.