Exchanging a Lamp
1
Press the release buttons and remove the front cover to 
gain access to the flow cell area. 
2
Disconnect lamp from the connector, unscrew the 
Vis-lamp (left) and/or UV-lamp (right) and remove the 
lamp. Do not touch the glass bulb with your fingers.
3
When replacing the Vis-lamp, assure that the Vis-lamp is 
inserted as shown (flat edge towards the deuterium 
lamp).
4
Insert the lamp. Fix the screws and reconnect the lamp 
to connector.