EasyManuals Logo

Agilent Technologies 1260 Infinity G1315C User Manual

Agilent Technologies 1260 Infinity G1315C
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #110 background imageLoading...
Page #110 background image
110 Agilent 1260 Infinity DAD and MWD User Manual
5
How to optimize the Detector
Optimizing for Sensitivity, Selectivity, Linearity and Dispersion
Slit Width
The detector has a variable slit at the entrance of the spectrograph. This is an
effective tool to adapt the detector to changing demand of different analytical
problems.
A narrow slit provides spectral resolution for analytes with very fine
structures in the absorbance spectrum. An example of such a spectrum is
benzene. The five main absorbance bands (fingers) are only 2.5 nm wide and
just 6 nm apart from each other.
Figure 38 Benzene at 1, 4 and 16 nm slit width (principle)
A wide slit uses more of the light shining through the flow cell. This gives
lower baseline noise as shown in Figure 39 on page 111.
QP
QP
QP

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315C and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315C Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals