140 Agilent 1260 Infinity DAD and MWD User Manual
7
Error Information
Detector Error Messages
Heater at fan assembly failed
Error ID: 1073
Every time the deuterium lamp or the tungsten lamp (DAD only) is switched
on or off a heater self-test is performed. If the test fails an error event is
created. As a result the temperature control is switched off.
Heater Power At Limit
Error ID: 1074
The available power of the heater reached either the upper or lower limit. This
event is sent only once per run. The parameter determines which limit has
been hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable cause Suggested actions
1
Defective connector or cable.
Please contact your Agilent service
representative.
2
Defective heater.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Excessive ambient temperature change.
Wait until temperature control equilibrates.