82 Agilent 1260 Infinity DAD and MWD User Manual
4
Using the Detector
Special Settings of the Detector
Automated Run Recovery in case of temporary communication
failures
Tabl e 1 2 Automated Run Recovery in case of temporary communication failures
Situation Reaction On ChemStation
All OK Run in progress - Data Analysis
Run / Rawdata
Elapsed run time is running
Spectra counter is running
Data are stored on PC and on card
LAN breaks Run in progress - Data Analysis
Run / Rawdata
Error Power Fail
Elapsed run time stops
Spectra counter stops
Data continues to be stored on card
LAN recovers Run in progress - Data Analysis
Run / Rawdata
Error Power Fail cleared
Elapsed run time continues at actual time
Data continues to be stored on PC and on card
ChemStation tries already to add missing data (depends on the
data load).