172 Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Exchanging a Flow Cell
Exchanging a Flow Cell
For bio-inert modules use bio-inert parts only!
When If an application needs a different type of flow cell or the flow cell needs repair.
Tools required p/n Description
Wrench, 1/4 inch
for capillary connections
OR 5043-0915 Fitting mounting tool
for bio-inert capillaries
Parts required # p/n Description
1 G1315-60022 Standard flow cell, 10 mm, 13 µL, 120 bar (12 MPa)
1 G1315-60025 Semi-micro flow cell, 6 mm, 5 µL, 120 bar (12 MPa)
1 G1315-60024 Micro flow cell, 3 mm, 2 µL, 120 bar (12 MPa)
1 G1315-60015 High pressure flow cell, 6 mm, 1.7 µL, 400 bar (40 MPa)
1 Nano flow cell, refer to “Nano Flow Cell - Replacing or Cleaning” on
page 192
1 G5615-60022 Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar (12 MPa) for
MWD/DAD, includes Capillary Kit Flow Cells BIO (p/n G5615-68755)
Preparations Turn th e lamp(s) off.
Remove the front cover.
CAUTION
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading
to sample degradation and contamination.
For bio-inert applications, always use dedicated bio-inert parts, which can be
identified by the bio-inert symbol or other markers described in this manual.
Do not mix bio-inert and non-inert modules or parts in a bio-inert system.