EasyManua.ls Logo

Agilent Technologies 1260 Infinity G1315C - Test Evaluation

Agilent Technologies 1260 Infinity G1315C
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity DAD and MWD User Manual 147
Test Functions
8
Filter Test
Test Evaluation
Filter Test Failed
Test Failed
Holmium Oxide Maximum out of Limits
Probable cause Suggested actions
1
Filter assembly (lever and filter) not
installed.
Install the filter assembly.
2
Defective filter motor.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Holmium oxide filter not installed.
Install the holmium oxide filter.
2
Dirty or contaminated filter.
Exchange the holmium oxide filter.

Table of Contents

Related product manuals