Agilent 1260 Infinity DAD and MWD User Manual 141
Error Information
7
Detector Error Messages
DSP Not Running
This error message comes up when the communication between the optical
unit and the main board has a problem.
No Run Data Available In Device
In a very rare case the capacity of the CompactFlash Card is not sufficient.
This could happen for example when the interrupt of LAN communication
takes longer and the detector uses special settings (e.g full data rate at 80 Hz
plus full spectra plus all signals) during data buffering.
Probable cause Suggested actions
1
Random communication error.
Switch the detector off and on again at the
power switch. If the error reoccurs:
Please contact your Agilent service
representative.
2
Defective detector main board.
Please contact your Agilent service
representative.
3
Defective PDA/optical unit.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
CompactFlash Card is full.
Correct communication problem.
Reduce data rate.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315C and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315C Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals