262 Agilent 1260 Infinity DAD and MWD User Manual
13
LAN Configuration
What you have to do first
What you have to do first
The module has an on-board LAN communication interface.
1 Note the MAC (Media Access Control) address for further reference. The
MAC or hardware address of the LAN interfaces is a world wide unique
identifier. No other network device will have the same hardware address.
The MAC address can be found on a label at the rear of the module
underneath the configuration switch (see Figure 85 on page 262).
Figure 84 MAC-Label
2 Connect the instrument's LAN interface (see Figure 85 on page 262) to
the PC network card using a crossover network cable (point-to-point) or
a hub or switch using a standard LAN cable.
Figure 85 Location of LAN interface and MAC label
3DUWQXPEHURIWKHGHWHFWRUPDLQERDUG
5HYLVLRQ&RGH9HQGRU<HDUDQG:HHNRIDVVHPEO\
0$&DGGUHVV
&RXQWU\RI2ULJLQ