Agilent 1260 Infinity DAD and MWD User Manual 201
Maintenance
9
Replacing Leak Handling System Parts
Replacing Leak Handling System Parts
1 Pull the leak funnel out of the leak funnel holder.
2 Pull out the leak funnel with the tubing.
3 Insert the leak funnel with the tubing in its position.
4 Insert the leak funnel into the leak funnel holder.
Figure 62 Replacing Leak Handling System Parts
5 Replace the front cover.
When If the parts are corroded or broken
Tools required None
Parts required # p/n Description
1 5041-8388 Leak funnel
1 5041-8389 Leak funnel holder
1 5062-2463 Corrugated tubing, PP, 6.5 mm id, 5 m
Preparations Remove the front cover.