Agilent 1260 Infinity DAD and MWD User Manual 171
Maintenance
9
Exchanging a Lamp
5
Replace the front cover.
Next Steps:
6
Reset the lamp counter as described in the user
interface documentation (lamps with I.D. tag cannot be
reset).
7
Turn the lamp on and give the lamp 10 minutes to warm
up.
8
Perform a “Wavelength Verification and Calibration” on
page 160 or a “Holmium Oxide Test” on page 151 to
check the correct positioning of the UV-lamp.
9
Perform an “Intensity Teston page 148.