EasyManuals Logo

Agilent Technologies 1260 Infinity G1315C User Manual

Agilent Technologies 1260 Infinity G1315C
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
102 Agilent 1260 Infinity DAD and MWD User Manual
5
How to optimize the Detector
Optimization Overview
Figure 32 Choosing a Flow Cell in HPLC
5 Setting the slit width
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance bands
and for high concentrations.
Use a wide slit (e.g. 16 nm) to detect very low concentrations.
Optimizing spectral acquisition (DAD only)
Select spectra acquisition mode according to your needs (see Tab l e 13 on
page 87).
Set the spectral wavelength range (for colorless samples 190...400 nm is
sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if high
resolution of spectra with fine structure is wanted.
spectral resolution, sensitivity and
linearity.
Tabl e 1 5 Optimization Overview
Parameter Impact
FP
FP
PLQ
PLQ
PLQ
PLQ
,QWHUQDOFROXPQGLDPHWHU PP PP PP PP
+LJKSUHVVXUH
IORZFHOOIRU
SUHVVXUHV
DERYHEDU
7\SLFDOFROXPQ
OHQJWK
7\SLFDOSHDN
ZLGWK
5HFRPPHQGHGIORZFHOO
7 FP
! FP
7\SLFDOIORZ
UDWH

POPLQ

POPLQ

POPLQ

POPLQ

POPLQ
6WDQGDUG
IORZFHOO
6HPLPLFUR
IORZFHOO
0LFURRU
6HPLQDQR

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity G1315C and is the answer not in the manual?

Agilent Technologies 1260 Infinity G1315C Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity G1315C
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals