102 Agilent 1260 Infinity DAD and MWD User Manual
5
How to optimize the Detector
Optimization Overview
Figure 32 Choosing a Flow Cell in HPLC
5 Setting the slit width
Use 4 nm slit for normal applications.
Use narrow slit (e.g 1 nm) if your analytes have narrow absorbance bands
and for high concentrations.
Use a wide slit (e.g. 16 nm) to detect very low concentrations.
Optimizing spectral acquisition (DAD only)
Select spectra acquisition mode according to your needs (see Tab l e 13 on
page 87).
Set the spectral wavelength range (for colorless samples 190...400 nm is
sufficient).
Set step to 4 nm for normal use; set small step (and slit width) if high
resolution of spectra with fine structure is wanted.
spectral resolution, sensitivity and
linearity.
Tabl e 1 5 Optimization Overview
Parameter Impact
FP
FP
PLQ
PLQ
PLQ
PLQ