EasyManua.ls Logo

Agilent Technologies 1260 Infinity G1315C - Page 175

Agilent Technologies 1260 Infinity G1315C
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity DAD and MWD User Manual 175
Maintenance
9
Exchanging a Flow Cell
7
Insert the flow cell capillaries into the union holder (top
is inlet, bottom is outlet). Tighten the thumb screw and
Reconnect the waste tubing (bottom) to the union.
To check for leaks, establish a flow and observe the
flow cell (outside of the cell compartment) and all
capillary connections.
Next Steps:
8
Perform a “Wavelength Verification and Calibration” on
page 160 or a “Holmium Oxide Test” on page 151 to
check the correct positioning of the flow cell.
9
Replace the front cover.
LQOHW
RXWOHW
ZDVWH
WKXPEVFUHZ

Table of Contents

Related product manuals