Agilent 1260 Infinity DAD and MWD User Manual 179
Maintenance
9
Maintenance of Standard, Semi-Micro or Micro Flow Cell
7
Press the window assembly onto the new or cleaned
quartz window.
8
Insert the window assembly [1] into the cell body.
Next Steps:
9
Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.
10
Reconnect the capillaries, see “Exchanging a Flow Cell” on page 172.
11
Perform a leak test.
12
Insert the flow cell.
13
Replace the front cover
14
Perform a “Wavelength Verification and Calibration” on page 160 or a “Holmium Oxide Test” on page 151 to check the
correct positioning of the flow cell.