EasyManua.ls Logo

Agilent Technologies 1260 Infinity G1315C - Page 194

Agilent Technologies 1260 Infinity G1315C
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
194 Agilent 1260 Infinity DAD and MWD User Manual
9
Maintenance
Nano Flow Cell - Replacing or Cleaning
5
The quartz body and the cell seal assembly can be
separated for cleaning purpose.
6
This figure shows the correct holding of the quartz body
and the cell seal assembly.
7
Replace the cell seal assembly onto the quartz body.
Always use a new seal assembly to exclude damage
during disassembling.
8
Slide the quartz body completely into the cell body to the
front stop (use for example a toothpick).

Table of Contents

Related product manuals