Agilent 1260 Infinity DAD and MWD User Manual 193
Maintenance
9
Nano Flow Cell - Replacing or Cleaning
1
Disconnect the capillaries from the capillary holder and
remove the flow cell.
2
Unscrew the cell body from the holder.
3
Unscrew the capillaries from the flow cell. DO NOT use
the adapter at this time!
4
Using for example a toothpick, press on the plastic part
and slide the quartz body out of the cell housing.