Agilent 1260 Infinity DAD and MWD User Manual 199
Maintenance
9
Cleaning or Exchanging the Holmium Oxide Filter
5
Replace the flow cell cover and fix the six screws.
Next Steps:
6
Perform a holmium oxide test, see “Holmium Oxide
Test on page 151 to check the proper function of the
holmium oxide filter.
7
Insert the flow cell, see “Exchanging a Flow Cell” on
page 172.
8
Replace the front cover.
9
Turn on the flow.