Agilent 1260 Infinity DAD and MWD User Manual 29
Site Requirements and Specifications
2 
Performance Specifications
Local Control Agilent Instant Pilot (G4208A) For 1260 systems:
• B.02.11 or above
For other systems:
• B.02.09 or above
Analog outputs Recorder/integrator: 100 mV or 1 V, 
output range 0.001 – 2 AU, two outputs
Communications Controller-area network (CAN), 
RS-232C, APG Remote: ready, start, 
stop and shut-down signals, LAN
Safety and 
maintenance
Extensive diagnostics, error detection 
and display (through control module 
and ChemStation), leak detection, safe 
leak handling, leak output signal for 
shutdown of pumping system. Low 
voltages in major maintenance areas.
GLP features RFID for electronics records of flow cell 
and UV lamp conditions (path length, 
volume, product number, serial number, 
test passed, usage)
Early maintenance feedback (EMF) for 
continuous tracking of instrument 
usage in terms of lamp burn time with 
user-setable limits and feedback 
messages. Electronic records of 
maintenance and errors. Verification of 
wavelength accuracy with built-in 
holmium oxide filter.
Housing All materials recyclable.
Others Electronic temperature control (ETC) for 
the complete optical unit
Tabl e 4 Performance Specifications G1315D
Type Specification Comments