32 Agilent 1260 Infinity DAD and MWD User Manual
2
Site Requirements and Specifications  
Performance Specifications
Control and data 
evaluation
Agilent ChemStation for LC (32-bit) For 1260 systems:
• Revision B.04.02 DSP2 or 
above
For 1100/1200 systems:
• Revision B.01.03 or above
Local Control Agilent Instant Pilot (G4208A) For 1260 systems:
• B.02.11 or above
For other systems:
• B.02.09 or above
Analog outputs Recorder/integrator: 100 mV or 1 V, 
output range 0.001 – 2 AU, two 
outputs
Communications Controller-area network (CAN), 
RS-232C, APG Remote: ready, start, 
stop and shut-down signals, LAN
Safety and 
maintenance
Extensive diagnostics, error detection 
and display (through control module 
and ChemStation), leak detection, 
safe leak handling, leak output signal 
for shutdown of pumping system. 
Low voltages in major maintenance 
areas.
GLP features RFID for electronics records of flow 
cell and UV lamp conditions (path 
length, volume, product number, 
serial number, test passed, usage)
Early maintenance feedback (EMF) 
for continuous tracking of instrument 
usage in terms of lamp burn time 
with user-setable limits and feedback 
messages. Electronic records of 
maintenance and errors. Verification 
of wavelength accuracy with built-in 
holmium oxide filter.
Housing All materials recyclable.
Others Electronic temperature control (ETC) 
for the complete optical unit
Tabl e 5 Performance Specifications G1365C