58 Agilent 1260 Infinity DAD and MWD User Manual
3
Installing the Module  
Installing Capillaries
5 When using UHP-FF fittings with bioinert capillaries, do not try to remove 
fittings from these capillaries. Bio-inert capillaries are using a PEEK front 
end, which may expand under pressure especially when being in contact 
with some organic solvents. If a fitting is moved across an expanded PEEK 
end, there is a risk of damaging the capillary by ripping off its end. Before 
re-installing such capillaries, push the ferrule towards the rear site for a 
small distance.
Figure 13 Capillary fitting