Chapter 3 Features and Functions
Remote Interface Configuration
3
Baud Rate Selection (RS-232)
You can select one of several baud rates for RS-232 operation.
When shipped from the factory, the baud rate is set to 57,600 baud.
You can set the baud rate from the front panel only.
Select one of the following: 300, 600, 1200, 2400, 4800, 9600, 19200,
38400, 57600 (factory setting), or 115200 baud.
The baud rate selection is stored in non-volatile memory and does not
change when power has been off or after a remote interface reset.
Front-Panel Operation: Press and then select the Baud Rate
softkey from the “I/O” menu.
Parity Selection (RS-232)
You can select the parity for RS-232 operation. When shipped from the
factory, the instrument is configured for no parity with 8 data bits.
You can set the parity from the front panel only.
Select one of the following: None (8 data bits), Even (7 data bits),
or Odd (7 data bits). When you set the parity, you are also setting the
number of data bits.
The parity selection is stored in non-volatile memory and does not
change when power has been off or after a remote interface reset.
Front-Panel Operation: Press and then select the Parity/# Bits
softkey from the “I/O” menu.

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals