Chapter 3 Features and Functions
Burst Mode
3
Burst Period
The burst period defines time from the start of one burst to the start of
the next burst. Used in the internal triggered burst mode only.
Keep in mind that burst period is different than the “waveform frequency”
which specifies the frequency of the bursted signal.
Burst period: 1 µs to 500 seconds. The default is 10 ms.
The burst period setting is used only when Internal triggering is
enabled. The burst period is ignored when manual or external
triggering is enabled (or when the gated burst mode is selected).
•It is not possible to specify a burst period which is too short for the
function generator to output with the specified burst count and
frequency (see below). If the burst period is too short, the function
generator will automatically adjust it as needed to continuously
re-trigger the burst.
Burst Period > + 200 ns
Front-Panel Operation
: To set the burst period, press the
Burst Period
softkey and then use the knob or numeric keypad to enter the period.
Remote Interface Operation:
BURSt:INTernal:PERiod {<seconds>|MINimum|MAXimum}
Burst Count
Waveform Frequency
-----------------------------------------------------------

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals