Chapter 5 Error Messages
Self-Test Errors
4
5
Self-Test Errors
The following errors indicate failures that may occur during a self-test.
Refer to the Agilent 33250A Service Guide for more information.
601 Self-test failed; system logic
This error indicates that the main CPU (U202) cannot communicate with
the main logic FPGA (U302).
602 Self-test failed; dsp
This error indicates that the main CPU (U202) cannot communicate with
the DSP (U506).
603 Self-test failed; waveform logic
This error indicates that the main CPU (U202) cannot communicate with
the waveform logic FPGA (U1201).
604 Self-test failed; even waveform memory bank
This error indicates that either the “even” waveform memory (U1304) or
the waveform logic FPGA circuitry (U1301, U1302, U1306) has failed.
605 Self-test failed; odd waveform memory bank
This error indicates that either the “odd” waveform memory (U1305) or
the waveform logic FPGA circuitry (U1302, U1303, U1307) has failed.
606 Self-test failed; cross-isolation interface
This error indicates that the I/O processor (U105) has either timed out or
failed its self-test.
607 to 614 607: Self-test failed; ground
608: Self-test failed; +16V supply
609: Self-test failed; +12V supply
610: Self-test failed; +5V supply
611: Self-test failed; +3.3V supply
612: Self-test failed; -2.1V supply
613: Self-test failed; -5.2V supply
614: Self-test failed; -16V supply
These errors indicate that the internal ADC has detected a power supply
with a voltage measurement outside of the expected range.