Error Messages
Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error that was stored. Errors are cleared as you
read them. The function generator beeps once each time an error
is generated (unless you have disabled the beeper).
If more than 20 errors have occurred, the last error stored in the
queue (the most recent error) is replaced with “Queue overflow”.
No additional errors are stored until you remove errors from the
queue. If no errors have occurred when you read the error queue,
the function generator responds with “No error”.
The error queue is cleared by the *CLS (clear status) command or
when the power is cycled. The errors are also cleared when you read
the error queue. The error queue
is not
cleared by an instrument reset
(*RST command).
Front-Panel Operation: Press and then select the topic entitled
View the remote command error queue” (topic number 2). Then press
the SELECT softkey to view the errors in the error queue. As shown
below, the first error in the list (i.e., the error at the top of the list) is
the first error that was generated.
Remote Interface Operation:
SYSTem:ERRor? Reads one error from the error queue
Errors have the following format (the error string may contain up to
255 characters).
-113,"Undefined header"