EasyManuals Logo

Agilent Technologies 33250A User Manual

Agilent Technologies 33250A
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
158
Chapter 4 Remote Interface Reference
Output Configuration Commands
4
VOLTage:OFFSet {<offset>|MINimum|MAXimum}
VOLTage:OFFSet? [MINimum|MAXimum]
Set the dc offset voltage. The default offset is 0 volts for all functions.
MIN selects the most negative dc offset voltage for the selected function
and amplitude. MAX selects the largest dc offset for the selected function
and amplitude. The :OFFS? query returns the offset voltage for the
function currently selected.
Limits Due to Amplitude
: The relationship between offset voltage and
output amplitude is shown below. Vmax is the maximum peak voltage
for the selected output termination (5 volts for a 50 load or 10 volts
for a high-impedance load).
|Voffset| <
Vmax
If the specified offset voltage is not valid, the function generator will
automatically adjust it to the maximum dc voltage allowed with the
amplitude specified. From the remote interface, a “Data out of range”
error will be generated and the offset will be adjusted as described.
Limits Due to Output Termination: The offset limits are determined
by the current output termination setting. For example, if you set the
offset to 100 mVdc and then change the output termination from
50 ohms to “high impedance”, the offset voltage displayed on the
function generator’s front-panel will double to 200 mVdc (and no error
will be generated). If you change from “high impedance” to 50 ohms,
the displayed offset will drop in half. See the OUTP:LOAD command on
page 163 for more information.
Arbitrary Waveform Limitations: For arbitrary waveforms, the
maximum
offset and amplitude will be limited if the waveform data
points do not
span the
full range of the output DAC (Digital-to-Analog
Converter).
For example, the built-in “Sinc” waveform does not use
the full range of values between ±1 and therefore its maximum offset
is limited to 4.95 volts (into 50 ohms).
Vpp
2
-----------

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish