212
Chapter 4 Remote Interface Reference
State Storage Commands
4
MEMory:STATe:RECall:AUTO {OFF|ON}
MEMory:STATe:RECall:AUTO?
Disable or enable the automatic recall of the power-down state from
storage location “0” when power is turned on. Select “ON” to automatically
recall the power-down state when power is turned on. Select “OFF”
(default) to issue a reset (*RST command) when power is turned on
(state “0” is not automatically recalled). The :AUTO? query returns
“0” (OFF) or “1” (ON).
MEMory:STATe:VALid? {0|1|2|3|4}
Query the specified storage location to determine if a valid state is
currently stored in that location. You can use this command before
sending the *RCL command to determine if a state has been previously
stored in this location. Returns “0” if no state has been stored or if it has
been deleted. Returns “1” if a valid state is stored in the specified location.
MEMory:NSTates?
Query the total number of memory locations available for state storage.
Always returns “5” (memory location “0” is included).