Chapter 4 Remote Interface Reference
The SCPI Status System
4
The Status Byte condition register is cleared when:
You execute the *CLS (clear status) command.
You read the event register from one of the other register groups
(only the corresponding bits are cleared in the condition register).
The Status Byte enable register is cleared when:
You execute the *SRE 0 command.
You turn the power on and have previously configured the function
generator to clear the enable register using the *PSC 1 command.
Note that the enable register will not be cleared at power-on if you
have configured the function generator using the *PSC 0 command.