ICP Supply
Measurement Lower
Limit
Upper Limit Measured Value Pass/Fail
Ch 1 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 2 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 31 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 4 Open Circuit Voltage 26 Vdc 32 Vdc Vdc
Ch 1 Current 2.75 mA 5.75 mA mA
Ch 2 Current 2.75 mA 5.75 mA mA
Ch 3 Current 2.75 mA 5.75 mA mA
Ch 4 Current 2.75 mA 5.75 mA mA
Source Amplitude Accuracy
Measurement Lower
Limit
(%)
Upper Limit
(%)
Measured Value
(%)
Pass/Fail
1 kHz, 0.1 Vpk -4 4
1 kHz, 3.0 Vpk -4 4
1 kHz, 5.0 Vpk -4 4
Source Output Resistance
Measurement Lower
Limit
Upper Limit
(ohm)
Measured Value
(ohm)
Pass/Fail
Resistance 5
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Agilent 35670A Verifying Specifications
Performance Test Record - Four Channel
19 of 20