EasyManuals Logo

Agilent Technologies 35670A Installation Guide

Agilent Technologies 35670A
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #156 background imageLoading...
Page #156 background image
Source Amplitude Accuracy
Measurement Lower
Limit
(%)
Upper Limit
(%)
Measured Value
(%)
Pass/Fail
1 kHz, 0.1 Vpk -4 4
1 kHz, 3.0 Vpk -4 4
1 kHz, 5.0 Vpk -4 4
Source Flatness
Measurement Lower
Limit
(dB)
Upper Limit
(dB)
Measured Value
(dB)
Pass/Fail
12.8 kHz -1 1
25.6 kHz -1 1
51.2 kHz -1 1
102.4 kHz -1 1
Source Distortion
Measurement Lower
Limit
Upper Limit
(dBc)
Measured Value
(dBc)
Pass/Fail
12.8 kHz -60
51.2 kHz -40
102.4 kHz -40
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Verifying Specifications Agilent 35670A
Operation Verification Test Record - Two Channel
10 of 10