5 Operating in EI Mode
To change the network settings of the TQ
98 7000/7010 Series TQ GC/MS Operating Manual
3 Click Edit NetConfig, and go to Edit NetConfig (QQQ network configuration).
(See Figure 27.)
4 Enter the GC IP Address in the DCOMM Partner IP Address if your GC
supports DCOMM. Your GC must be an Agilent 7890B or 7890A+ with the
latest firmware for DCOMM support.
5 Enter the IP Address, Gateway IPA, and Subnet Mask for the TQ.
6 Click Submit to assign this IP address to the TQ.
Agilent recommends attaching a label to the TQ recording these NetConfig
parameters.
7 Close the browser page when you are done.
8 Run the Agilent GC/MS Instrument Configuration program. To configure the
TQ for DCOMM you must first configure the GC before configuring the TQ in
Figure 27. WUI Edit NetConfig