Chromatographic Symptoms 2
Agilent 7697A Troubleshooting 19
Retention Times Not Repeatable
Check the GC. See the GC’s troubleshooting information.
GC problems can include:
Leaks in the inlet, including septum
Leaks around the inlet liner, if applicable
Gas supply pressure
Wrong liner type used, if applicable
First run effects (has the GC stabilized?)
Check the HS sequence log, event log, and status displays
for messages. Especially check the sequence log for
method deviations and dynamic leak check failures.
If using HS carrier gas control, confirm that the current
HS column configuration matches the column actually
installed in the GC.
Check for leaks in the transfer line to the GC inlet,
including the connection to the inlet.
Check for leaks in the other parts of the sampling system
(probe, six port valve, and lines and fittings between).
Enable dynamic leak checking. This feature verifies that
each sample vial is leak- free.
If not enabled, check the sample vial for cracks or
other leaks. See the Operation Guide.
If using dynamic leak checking, make sure the entered
leak rate is appropriate for the sample and its
preparation technique.

Table of Contents

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals