68 Agilent 7697A Troubleshooting
5 Vial Handling
Barcode Reader Errors
If using the optional barcode reader, a barcode error
indicates that the barcode reader could not read the barcode
as expected.
Check the vial barcode label positioning, sizing, and type.
See the Operating guide.
Barcode label not heat resistant.
Check vial locations. Was this vial the correct vial?
Check the barcode label. Is the label smudged or illegible?
Is the label too glossy? If possible, try reading the
barcode using another device.
Check the HS sequence log, event log, and status displays
for messages.
Check the method. Does the barcode type specified in the
method match the barcode applied to the vial?
If the barcode failed checksum or its value did not match
the expected value, check vial locations. Check barcode
type used.
If the problem persists, try test vials with new labels. Be
sure each vial meets the labeling requirements.