Vial Handling 5
Agilent 7697A Troubleshooting 63
Carousel Errors
A carousel error indicates that the carousel could not
function properly. For a 12 vial model, this is the vial tray.
For the 111 vial model, the carousel resides in the vial oven.
Possible causes include:
Broken vial in the oven
Obstruction in the shutter. (111 vial model)
Other obstruction in the oven
Carousel stepper motor issue
Belt issue
Before proceeding, cool down the vial oven (111 vial model)
Check the HS sequence log, event log, and status displays
for messages. (Was this error caused by another problem?)
Check the oven shutter. Is it open or closed?
Check the carousel/tray for broken glass or other
obstructions. See the procedure for cleaning the 12 vial
oven tray or vial oven in the Maintenance manual.
If the problem remains, contact Agilent for service.
WARNING
The oven and contents in it may be hot enough to cause burns.
Cool the oven to a safe handling temperature before proceeding,
or wear heat-resistant gloves.
WARNING
If there is a broken vial in the oven, the oven will contain sharp,
broken glass as well the contents of any sample material.

Table of Contents

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals