EasyManua.ls Logo

Agilent Technologies 7697A - What Is a Sequence?

Agilent Technologies 7697A
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 Agilent 7697A Troubleshooting
5 Vial Handling
Sample Probe Lifter Errors
The sample probe lifter raises the sample vial onto the
sampling probe and lowers it back into the rotating tray. A
sample probe lifter error indicates that the lifter could not
function properly.
Check the HS sequence log, event log, and status displays
for messages. (Was this error caused by another problem?)
Check the carousel/tray for broken glass or other
obstructions. See the procedure for cleaning the 12 vial
oven tray or vial oven in the Maintenance manual.
Check vial size configuration.
Contact Agilent for service.

Table of Contents

Other manuals for Agilent Technologies 7697A

Related product manuals