42 Agilent 7697A Troubleshooting
4 Leaks
If not already done, check the GC. The HS leak test
cannot test for leaks in the GC.
If the test fails
If the test fails, the display provides:
The reading for the stage that failed (for example, leak
rate or flow rate).
Command lines to toggle related valves (for example, a
switching valve or the six port valve).
SV1 or SV2 or PV1 or PV2: Scroll to this line and press
[On/Yes] to turn on (energize) the valve, or [Off/No] to
turn it off.
Six-port valve: Scroll to this line and press [On/Yes] to
switch the valve to the load position, or [Off/No] to switch
it to the inject position.
A selection to exit the test.
Exit test?: Select this line then press [On/Yes] to abort the
test.
A failure code
See the sections below for troubleshooting information for
each stage of the test.