Communications 7
Agilent 7697A Troubleshooting 85
The GC Does Not Start After the HS Injects
If the HS makes an injection, but the GC does not start a
run, check the following:
Headspace sampler multiple headspace extraction mode
for this method. Press [Adv Function] and check the
Extraction Mode. If performing concentrating extractions,
the HS only starts a GC run after the final concentrating
injection.
APG remote cable connection.
The HS method APG Remote enable setting. If turned off,
the HS will not wait for GC readiness before injecting.
System Not Ready sequence action setting for the HS
method. If set to continue, the HS will inject regardless of
GC readiness state.
GC readiness. If the GC is not ready, it will not start.