Chromatographic Symptoms 2
Agilent 7697A Troubleshooting 21
Low Peak Area or Height (Low Sensitivity)
Check gas supply purity.
Check all trap indicators and dates.
Check the GC.
If possible, mount an automatic liquid sampler (ALS)
over the inlet and inject a sequence of internal or
calibration standard directly into the inlet. Check the
sensitivity. If acceptable, check the HS. If unacceptable,
the problem is in the GC.
If the problem is in the GC, refer to the GC’s
documentation for troubleshooting procedures.
Check sample vial size. (The HS cannot distinguish
between 20- mL and 22- mL vials.)
Check the configured sample loop size. If incorrect, the
peak areas may be reduced due to insufficient headspace
pressure for filling a bigger sample loop than configured.
Check the HS sequence log, event log, and status displays
for messages. Especially check the sequence log for
method deviations and dynamic leak check failures.
Check that the transfer line is correctly installed at both
ends.
Check for leaks at the six port valve and within the rest
of the sampling system (sample probe, its connections to
the six port valve).

Table of Contents

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals