Leaks 4
Agilent 7697A Troubleshooting 39
1
Cool the HS sample loop, transfer line, and GC inlet and
column oven to prevent damage during the test.
2 Turn off the HS.
3 Set the regulator pressure to 415 kPa (60 psi).
4 Fully turn the regulator pressure adjustment knob
counterclockwise to shut the valve.
5 Wait 5 min. If there is a measurable drop in pressure,
there is a leak in the external connections. No drop in
pressure indicates that the external connections are not
leaking.