288 Chapter6
Troubleshooting the RF Section
6b. 8592L/94L, 8594Q, and 8593E/94E/95E/96E
Table 6-4 Power Levels at Measurement Points
Measurement
Point
Measurement
Frequency
Power Level Range *
Band 0
92L, 93E,
94E, 94L,
94Q, 95E,
96E
Band 1
92L, 93E,
95E, 96E
Band 2
92L, 93E,
96E
Band 3
92L, 93E
Band 4
92L,93E
A Same as input
frequency
30 to 33
dBm
————
B Same as input
frequency
14 to 19
dBm
14 to 19
dBm
14 to 20
dBm
15 to 23
dBm
C 321.4 MHz 10 to 16 dB
below B †
18 to 26 dB
below B †
21 to 31 dB
below B †
25 to 35 dB
below B †
D 3.9214 GHz 8 to 12 dB
below A †
————
E 321.4 MHz 5 to 11 dB
below D †
1 to 6 dB
below C †
1 to 6 dB
below C †
1 to 6 dB
below C †
1 to 6 dB
below C †
* A frequency-selective measuring device, such as another analyzer, is recommended for making these
measurements. Broadband measuring devices, such as power meters, will give erroneous results due to the
presence of other, higher-amplitude signals.
† The power-level range is relative to the actual measurement taken at the measurement point indicated.