70 Chapter2
Making Adjustments
9. CAL AMPTD Adjustment Routine
9. CAL AMPTD Adjustment Routine
This adjustment applies to:
All 8590 E-Series and L-Series spectrum analyzers
8591C cable TV analyzers
8594Q QAM analyzers
The
CAL AMPTD softkey accesses an internal self-adjustment routine.
The following adjustments are automatically performed by
CAL AMPTD.
The reference level is calibrated by adjusting the gain of the IF
section.
The 3 dB resolution bandwidths are adjusted.
Bandwidth amplitude errors are determined. Errors are corrected
with video offsets.
Step-gain and input-attenuator errors are determined. Errors are
corrected with video offsets.
Log fidelity is checked in 1 dB steps. Errors are corrected with video
offsets.
Frequency accuracy is tested and adjusted by using frequency
offsets.
Equipment
Cable, BNC, 20 cm (9 in)
Adapter, Type N (m) to BNC (f)
Additional Equipment for Option 026
Adapter, Type APC-3.5 (f) to APC-3.5 (f)
Additional Equipment for 75 Input
Cable, BNC, 75 , 30 cm (12 in)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8590L and is the answer not in the manual?

Agilent Technologies 8590L Specifications

General IconGeneral
BrandAgilent Technologies
Model8590L
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals