Chapter 14 617
Analyzer Messages
Interpreting Analyzer Messages
Before removing the A16 processor/video assembly, check all assemblies
that are identified in Table 14-1.
1. Turn the instrument power off and remove each assembly, one at a
time.
2. Turn the instrument on.
If the ADC ground fault remains, the instrument preset at
power-up displays the error message.
If the error message goes away, the last assembly removed is the
cause of the problem.
If the error message remains after the removal of all related
assemblies, suspect the A16 processor/video assembly.
An ADC ground failure can also be checked manually, using the service
diagnostic. Press CAL, More 1 of 4, More 2 of 4, Service Diag, More 1, GND
REF DETECTOR.