Chapter 14 637
Analyzer Messages
Interpreting Analyzer Messages
Related Assemblies
A7 analog interface, A16 processor/video assembly, A17 memory card,
A25 counter lock, assemblies installed in the card cage.
Troubleshooting Hints
Refer to Figure 9-1, IF/Control Overall Block Diagram, and Figure 9-2,
Figure 9-3, or Figure 9-4, RF Section Overall Block Diagram for your
analyzer, in Chapter 9 for an overview of the assemblies attached to
the I/O bus. Refer also to “Troubleshooting the A15 Motherboard
Assembly” in Chapter 5 for the location of specific I/O signal traces on
related assemblies.
If a failure code for an I/O bus address or data line occurs, remove
the related assemblies one at a time until the failure message
disappears. Be sure to turn the analyzer power off while removing
and installing each assembly.
If the failure message disappears, suspect the assembly that was
removed last.
If the failure message remains after all related assemblies have
been removed, suspect the A16 assembly.
If a A16 video RAM fail code occurs, the A16 assembly is defective.
File type incompatible (U)
Indicates that the selected file is not a display image file. The file name
for a display image file is always preceded by an “i.