644 Chapter14
Analyzer Messages
Interpreting Analyzer Messages
Memory Overflow: ERASE DLP MEM and reload (U)
This message indicates that too many user-defined items (functions,
variables, key definitions), or downloadable programs have been loaded
into analyzer memory.
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
Use DISPOSE USER MEM and then load the user-defined item or
downloadable program into analyzer memory.
Press the following analyzer keys.
CONFIG, More 1 of 3
DISPOSE USER MEM
If the above keys are not accessible perform the following steps.
1. Press the following analyzer keys.
DISPLAY
Change Title
Use the knob to select the letters DISPOSE ALL;, be sure to include
the semicolon (;).
2. Press the following analyzer keys.
CAL, More 1 of 4, More 2 of 4
SERVICE CAL
EXECUTE TITLE
If the analyzer still appears to be locked up, refer to “Isolating an RF,
LO, IF, or Video Problem” in Chapter 4 .
INVALID TRACE: _ _ _ (U)
The specified trace is invalid.
INVALID TRACE NAME: _ _ _ (U)
The specified trace name is not allowed. Use an underscore as the
second character in the trace name, or avoid beginning the trace name
with the following pairs of letters: LB, OA, OL, TA, TB, TR, MA, MF,
TS, OT, and DR.