Chapter 2 75
Making Adjustments
10. CAL YTF Adjustment Routine
7. Press the following keys to allow entry of the default mixer bias DAC
value for band 2.
DISPLAY
CHANGE TITLE
CLEAR
8. Enter CALMXRDATA 20,2; as a title entry to set the mixer bias value
to 20 for band 2.
Press the following keys to store the mixer bias value for band 2 in
nonvolatile memory.
CAL, More 1 of 4, More 2 of 4
SERVICE CAL
EXECUTE TITLE
9. Press the following keys to allow entry of the default mixer bias DAC
value for band 3.
DISPLAY
CHANGE TITLE
CLEAR
10.Enter CALMXRDATA 1300,3; as a title entry to set the mixer bias
value to 1300 for band 3.
11.Press the following keys to store the mixer bias value for band 3 in
nonvolatile memory.
CAL, More 1 of 4, More 2 of 4
SERVICE CAL
EXECUTE TITLE
12.Press the following keys to allow entry of the default mixer bias DAC
value for band 4.
DISPLAY
CHANGE TITLE
CLEAR
13.Enter CALMXRDATA 520,4; as a title entry to set the mixer bias
value to 520 for band 4.