46 1260 Infinity Binary LC - System User Guide
3 Optimization of the Agilent 1260 Infinity Binary LC
How to Configure the Optimum Delay Volume
The standard delay volume configuration is also the configuration which
provides direct method transferability from the Agilent 1100 and 1200
Series LC system to the 1260 Infinity Binary LC or vice versa. Figure
Figure 16 on page 46 shows two overlaid chromatograms of a method for
analysis of paracetamol and related impurities. The method was
transferred from an Agilent 1200 Series LC system to an Agilent 1260
Infinity Binary LC with the chromatographic conditions (column, mobile
phase, pump settings, injection volume, column temperature, detector
settings) left unchanged. As can be seen, a seamless method transfer is
possible.
Figure 16 Comparison Agilent 1200 Series and Agilent 1260 Infinity LC system on one
Configuration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 1260 Infinity and is the answer not in the manual?

Agilent Technologies Agilent 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals