1260 Infinity Binary LC - System User Guide 89
Quick Start Guide
5
Data Acquisition in Method and Run Control View
Tab le 9
Method parameters for first separation run - standard delay volume configuration
Module Parameter Setting
Pump Solvent A Water
Solvent B Acetonitrile
Flow rate 4.0 mL/min
Solvent
Composition
40 % A, 60 % B
Stop Time 1 min (1 min post time)
Autosampler Injection 1 µL
Column
Compartment
Column Column Eclipse Plus C18, 4.6 x 100 mm, 3.5 µm
(959961-902)
Column Poroshell 120 EC-C18, 3.0 x 50 mm, 2.7 µm
(699975-302)
Column Poroshell 120 EC-C18, 4.6 x 50 mm, 2.7 µm
(699975-902)
Te m p er a t u re 4 0 ° C
Detector Signal A Wavelength: 250 nm, Bandwidth: 4 nm
Reference Wavelength: 360 nm, Reference
Bandwidth: 100 nm